SPYDA BIBS

SPYDA BIBS

SPYDA BIBS

Product catalog summary
Introduction
The SPYDA BIBS (Built in Breathing System) is designed for use in hyperbaric chambers to supply pure oxygen during therapeutic treatments and as an emergency breathing source in case of contamination or fire.
Design and Functionality
This system features an 'overboard dump' style with both supply and exhaust hoses to prevent hazardous oxygen buildup. It is optimized for minimal breathing resistance at low rates and shallow depths, suitable for therapeutic decompression starting at 18m.
Robustness and Practical Design
The SPYDA BIBS is constructed from advanced polymers for durability, with an easily removable oral nasal for personal use or decontamination.
Testing and Compliance
The system complies with ABS standards, tested to 300MSW/1000FSW, and exceeds requirements for underwater vehicles and hyperbaric facilities. It offers excellent Work of Breathing at typical rates.
Use with 100% Oxygen
All components are cleaned according to the National Hyperbaric Centre’s procedures to prevent contamination and fire risks. Consultation with Spyda Systems Ltd or technical authorities is advised if in doubt.
Operating Requirements
The inlet pressure should be set at 8.5 bar above ambient pressure, with outlet hose suction not exceeding 2 Bar. A Back Pressure regulator is recommended beyond 20m external chamber pressure.
Maintenance
The BIBS mask should be part of a Planned Maintenance System (PMS), with guidance available in the IMCA D018 DESIGN/DASS document. Maintenance instructions, including chemical and biological procedures, are provided in the manual.
See more

Catalog excerpts

SPYDA BIBS-1

SPYDA BIBS A robust oxygen mask for use in dry hyperbaric enviroments The SPYDA BIBS (Built in Breathing System) is designed to supply pure oxygen during therapeutic treatment in a hyperbaric chamber. It can also be utilised to provide an emergency breathing gas source should the chamber environment become chemically contaminated or experience a fire. It is an ‘overboard dump’ style BIBS; it has both supply and exhaust hoses. Exhaled gas is piped to the outside of the chamber thus preventing the potentially hazardous build up of oxygen which occurs when single hose units are installed. An increase in oxygen level above 21% equivalent at atmospheric pressure (0.21 bar ppO2) within a hyperbaric chamber can be a fire risk, and can also be detrimental to the health of the patient/diver if long exposure is experienced. The main use of BIBS masks is for therapeutic decompression commencing at 18m. The Spyda unit has been optimised to offer minimum breathing resistance at relatively low rates of breathing and shallow depths associated with this procedure.

 Open the catalog to page 1
SPYDA BIBS-2

Robustness and Practical Design Users are now demanding equipment that can withstand the bumps, bangs and physical abuse from day to day operations, yet still deliver first rate performance. The advanced polymer used for the main body is extremely robust and will undoubtedly provide longevity in service. The oral nasal is easy to remove and replace. This permits divers to have their own oral nasal, or otherwise permit easy biological decontamination. Testing The Spyda BIBS have been extensively tested and fully comply with ABS (American Bureau of shipping) standards under their latest rules for...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.