APC-2000ALM

APC-2000ALM

APC-2000ALM

Product catalog summary
Overview: The document provides detailed technical specifications and operational guidelines for the APC-2000ALM Smart Pressure Transmitter, which utilizes the MODBUS RTU protocol. It includes information on various configurations, materials, and installation procedures.
Specifications:
  • Accuracy: ±0.075% (±0.1% for range no. 18)
  • Long-term Stability: Accuracy for 3 years or 2x accuracy for 5 years
  • Thermal Error: < ±0.05% (FSO) / 10°C
  • Thermal Compensation Range: -25 to 80°C
  • Materials: Wetted parts and diaphragm made of 316L stainless steel; casing available in aluminum or 316 stainless steel
  • Ingress Protection: IP66
  • Operating Temperature Range: -40 to 85°C (ambient), -40 to 75°C (Exd)
  • Medium Temperature Range: -40 to 120°C
Electrical Parameters:
  • Power Supply: 12 to 30 V DC
  • Current Consumption: < 80 mA
  • Output: MODBUS RTU
  • Resolution: 0.0007% of measuring range
  • Transmission Range: 1200m
  • Transmission Speed: 1200 to 115200 bps
  • Factory Settings: Transmission speed 9600 bps, parity check even, address 1
Installation and Application:
  • The transmitter can be mounted directly on the object. For steam or hot media, use a siphon or impulse tube.
  • Equipped with a configurable LCD display that can be rotated and adjusted for optimal viewing.
  • Designed to measure pressure, vacuum, and absolute pressure of gases, vapors, and liquids.
Operation Modes:
  • Operates in a four-wire system with RS485 transmission using the Modbus RTU protocol.
  • Allows setting of transmission parameters, network address, and resetting/calibrating the transmitter.
Ordering Procedure:
  • Model Code: APC-2000 with various options for display, housing material, and certificates.
  • Process Connections: Available in multiple thread types and materials, including options for gold-plated and Hastelloy diaphragms.
  • Electrical Connection: Options for packing gland and thread types.
  • Accessories: Includes mounting brackets and stainless steel plates.
See more

Catalog excerpts

APC-2000ALM-1

SMART PRESSURE TRANSMITTER APC-2000ALM with MODBUS RTU protocol ü ü ü ü Galvanic isolation of power supply and communication bus Accuracy 0.075% Aluminum or stainless steel housing with local display Explosion proof certificate ATEX and IECEx APC-2000ALM - Aluminum epoxy painted casing APC-2000ALM/SS - Stainless steel casing Rating plate Lock preventing the casing rotation 1/2’’NPT M type 1/2’’NPT male + internal thread G1/4’’ G1/2 type G1/2", Æ4 hole M type M20×1.5, Æ4 hole M20×1.5 packing gland Æ5…10 cable or 1/2 NPT F 1/2’’NPT F type internal thread 1/2-14NPT GP type G1/2", Æ12 hole P type M20×1.5, Æ12 hole GP type P type 1/2"NPT M type for HS version CG1 type G1" with flush diaphragm CG1/2 type G1/2" with flush diaphragm Version with direct or remote diaphragm

 Open the catalog to page 1
APC-2000ALM-2

Application Pressure transmitter APC-2000ALM is designed to measure pressure, vacuum and absolute pressure of gases, vapors and liquids. The measuring element is a piezoresistive silicon sensor separated from the medium by an isolating diaphragm and manometer liquid. The electrical signal from the sensor is converted into a digital value compatible with the Modbus RTU protocol. The housing is made of high-pressure cast aluminum alloy or acid-resistant steel with IP66 protection. The transmitter is equipped with a configurable liquid crystal display. The housing design allows the display to be...

 Open the catalog to page 2
APC-2000ALM-3

Ordering procedure Model APC-2000 Versions Certificates, options* * more than one option is available Nominal measuring range Description Smart pressure transmitter With display, Modbus output Stainless steel housing II 1/2G Ex ia/db IIC T5/T4 Ga/Gb II 2D Ex ia tb IIIC T100°C Db Packing gland available on I M2 Ex db ia I Mb (version with SS housing) request Ex ia/db IIC T5/T4 Ga/Gb IECEx Ex ia tb IIIC T100°C Db Ex db ia I Mb (version with SS housing) For oxygen service (sensor filled with Fluorolube fluid), only G1/2” connection Extended thermal compensation range -40 ÷ 80°C Protection class...

 Open the catalog to page 3

All APLISENS S.A. catalogs and brochures

  1. PEM-500

    4  Pages

  2. MS-100

    1  Page

  3. MS-100K

    1  Page

  4. PR - G

    2  Pages

  5. APR -2000G

    3  Pages

  6. APR-2000ALW/G

    3  Pages

  7. APM-2

    3  Pages

  8. APR-2000ALW

    7  Pages

  9. AS-dP

    1  Page

  10. PRE-28.MODBUS

    3  Pages

  11. PRE-28.SMART

    3  Pages

  12. APRE-2000

    3  Pages

  13. APCE-2000

    3  Pages

  14. AS

    1  Page

  15. PCE-28.MODBUS

    3  Pages

  16. PCE-28

    3  Pages

  17. PCE-28.SMART

    3  Pages

  18. APC-2000ALW

    5  Pages

  19. PEM-1000

    4  Pages

  20. RH

    3  Pages

  21. ERH-xx-20

    2  Pages

  22. ERH-11-20

    1  Page

  23. ERH-SMALL

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.